  • The testing machine host power source is not bright,the equipment can not move up and down.The testing machine host power source is not bright,the equipment can not move up and down.
     The host is electric,but the testing machine does not operate.

     Check the connection of the power line accessed testing machine whether is normal; Check the emergency stop switch whether is in the closed state;Check the fuse of socket whether has blown,if so take out the spare fuse to setup.

 

              The host is electric,but the testing machine does not operate.

            A)The power supply voltage is whether or not normal,if not please add UPS power source.
            B)The communication line of host and computer has a bad connection,put the plug tightly and locked.
            C)The communication serial port of computer has broken,please replace good communication serial port.
            D)  The limit switch is in working,transfer the limit  circlip to the appropriate location

              The sample has not been completely torn,peeled and tensiled,the test has stopped.


            A)The judging condition of test scheme for soft is not set correctly,modify it. 
            B)Testing force has reached the maximal force of testing machine,this time should saw the sample,in order to avoid damage to the transducer.